Signal Flow and Data Pipeline

ADC acquisition cycle (per 0.1 s tick)

  1. ADC.acquisition_loop() runs in its own QThread.
  2. self.pause(self.sampling_time) (default 0.1 s) — an abort-aware sleep.
  3. collect_data() reads N channels via aio.analog_read_volt(channel, datarate, gain). The PCA9554 mux is reconfigured only when the channel range demands it.
  4. Raw row appended to adc_values DataFrame; converted row appended to converted_values.
  5. If a plasma-current setpoint is non-zero: plasma_current_control() runs simple_pid.PID(0.3, 0.1, 0) against Ip and emits send_control_voltage → MCP4725.
  6. Every STEP ticks: send_processed_data_to_main_thread() emits data_ready([dataframe, device_name]). MainApp.on_worker_step routes to _adc_step, appends to self.datadict["ADC"], calls save_data (CSV append), triggers plot update.

Averaging at slow sampling

A period of one second or longer is not read once. From AVERAGE_FROM_SECONDS = 1.0 upwards (controlunit/devices/adc.py) the reader converts every INNER_SECONDS = 0.2 seconds through the period and records one row at the period's end whose raw voltages are the mean of those readings — fifty of them fill the rig's ten-second setting. Below the threshold nothing changed: one conversion per period, because fast sampling is for watching transients and the instant is the point. Slow sampling is an overnight or weekend log, where a single ~1 ms conversion records whatever noise sat on the line at that instant and the period's mean is the truer number (owner decision 2026-09-07).

The mean is taken over the raw voltages and converted afterwards, exactly as one reading is converted, so a row stays self-consistent: converting the raw column of the CSV reproduces the converted column beside it. The columns, the timestamp (the period's end) and the one-row-per-period shape are unchanged, so the CSV, the web ring and PIHTI Log see nothing new. The period is measured with time.monotonic() and each inner conversion is due at a fixed offset from its start, so the conversions' own time comes out of the waits and ten seconds of sampling still take ten seconds. Both the inner cadence and the period wait through DeviceThread.pause; an abort mid-period ends the loop at once and the half-measured period records nothing. The plasma-current PID runs once per recorded row, on the averaged value.

STEP batching

# controlunit/devices/adc.py
if step % (self.STEP - 1) == 0 and step != 0:
    self.send_processed_data_to_main_thread()
    step = 0
else:
    step += 1

STEP serves two purposes simultaneously: averages noisy ADC samples and amortises Qt signal-emission overhead. One parameter, two jobs.

STEP is set dynamically by sampling rate:

# controlunit/devices/device.py
def set_sampling_time(self, sampling_time):
    if sampling_time >= 0.9:  self.STEP = 1
    if sampling_time < 0.9:   self.STEP = 3
    if sampling_time < 0.1:   self.STEP = 5

Plasma current PID

Live. Uses simple_pid.

  • p=0.3, i=0.1, d=0, output_limits=(0, 4500) mV.
  • Setpoint from GUI; feedback from Hall-effect sensor on channel 0.
  • Actuator: MCP4725 DAC behind galvanic I²C isolator (Apr 2026).
  • baseline = 1000 mV — empirical minimum for plasma ignition (Kawabata-kun).

Membrane heater PID

Dormant. Code exists in MAX6675.temperature_control(). Hand-rolled: integral-clamped (if integral < -0.5: integral = 0), asymmetric (only positive e drives output). Gains Kp=3.5, Ki=0.06, Kd=0. Output is on-time fraction of a 10 ms cycle, software-PWMed against an SSR.

The measurement and PID have migrated to a Windows machine with NI hardware. Not returning to the Pi; if it returns at all, it will be a dedicated ESP32 unit reporting to an orchestrator.

GPIO sync signal

GPIO 26 emits a digital sync edge consumed by various external data loggers (QMS and others). The front-panel LED is a visual indicator only — not an isolation strategy, not opto-isolated triggering. The shared GPIO was always a wiring convenience.

QMS_signal in the CSV is a boolean column logging whether the sync trigger is currently active during each ADC row.

Safety stop semantics

# controlunit/main.py
def turn_off_voltages(self):
    self.workers["ADC"]["worker"].set_plasma_current.emit(0)
    self.workers["PlasmaCurrent"]["worker"].output_voltage_signal.emit(0)
    self._mfc_presets = {1: 0, 2: 0}
    self.update_current_values()
    self.workers["MFCs"]["worker"].output_voltage_signal.emit(1, 0)
    self.workers["MFCs"]["worker"].output_voltage_signal.emit(2, 0)

abort_all_threads → turn_off_voltages first, thread termination second. Plasma setpoint goes to zero before any thread dies.

Logging

Two parallel append-only paths:

  1. CSV at ~/work/cudata/cu_<YYYYMMDD_HHMMSS>.csv with a self-describing comment header. Header embeds enough channel metadata that an old CSV can be replayed without settings.yml.
  2. Event log at ~/work/cudata/controlunit.log.

Every ADC row carries commanded presets alongside measured signals: PresetV_mfc1, PresetV_mfc2, PresetV_cathode, IGmode, IGscale, IGmode_Pu2, IGscale_Pu2 (one mode and exponent pair per ionization gauge, named by the gauge's Mode Column and Scale Column in settings.yml), QMS_signal.