Logging the Kikusui over LAN

The cathode's own voltage and current are to be logged from the Kikusui PWR401L (PWR-01 series) over its LAN port, not from its analog monitor pins and not from a Hall sensor (owner decision 2026-09-11, "lan it is then"). The reasoning is in Cathode control and in the lab vault; this page is the how-to, with everything below taken from Kikusui's PWR-01 Interface Manual (IB035082, 2020) and the PWR-01 FAQ.

Why LAN

  • J1's monitor outputs are referenced to the supply's output negative, which floats at cathode potential; reading them would need a second isolation stage and one more analog path into the ADC board whose I²C bus already fails under an arc.
  • Ethernet is transformer-isolated by construction. Nothing electrical ties the supply to the Pi.
  • The supply's own metering is calibrated and answers in volts and amperes; no rerouting of 30 A through a sensor.

On the supply, once

  1. Connect a straight category-5 cable from the rear LAN port (it has a cover; the RS232C port beside it is the same shape, read the label) to the lab switch.
  2. Enable the interface: press CONFIG until CF40 shows, turn the CURRENT knob to ON.
  3. Choose how it gets an address: press CONFIG until CF60 shows, turn the VOLTAGE knob to CF61, and set with the CURRENT knob: 110 = DHCP on, Auto-IP on, manual off (the factory default; fine on the lab VLAN if the router gives leases); 001 = manual, then enter the four octets in CF62 to CF65 and the subnet mask as a bit count in CF66 (16 for 255.255.0.0, 24 for 255.255.255.0). The lab VLAN's addresses have been static for years, so a manual address beside the Pi's is the sensible choice.
  4. Apply: CF60: APPL restarts the interface, or turn the POWER switch off and on. Changes to CF61–CF66 take effect only then.
  5. Open http://<address>/ in a browser: the interface board serves a page with the instrument information, the network settings and a remote-control panel. Setting a password there is optional; CF60: LCI resets the interface if the password is forgotten.

Read-only logging (4.15.0)

The first stage records the supply while the operator drives discharges and bakes manually. Collect normal behaviour before changing the plasma PID or inventing filament-condition warning thresholds (owner decision 2026-09-14). There is no automatic ignition, and until 2026-09-15 this logger sent no command at all; it now sends the two output writes below, and only when a person presses for them (see The two writes).

On the Pi, create ~/.controlunit/kikusui.yml with its actual address:

host: 192.0.2.10  # example only: replace with the supply's numeric LAN address
port: 5025
interval_s: 0.5
timeout_s: 1.0
retry_s: 5.0

Configuration is read at acquisition Start. An absent file disables logging with one message; an invalid file reports the problem without preventing normal acquisition. The host is a numeric IPv4 or IPv6 address so DNS cannot hold up shutdown. interval_s accepts 0.1–60 s, timeout_s 0.1–5 s, and retry_s 1–60 s. The defaults record at about 2 Hz independently of the ADC's sampling period. Long polls reduce telemetry cadence, never cause catch-up bursts. For off-rig testing use dummy: true; rows say dummy and contain synthetic zero output. Dummy hardware refuses real LAN configuration.

Each run creates kikusui_YYYYMMDD_HHMMSS.csv beside its cu_YYYYMMDD_HHMMSS.csv. A header identifies the paired ADC file and schema controlunit-kikusui/v1. The columns are:

Column Meaning
date Receipt time in ISO format with the local UTC offset
logger_elapsed_s Monotonic time since this recorder started
query_ms Time for the poll, including reconnect/identity when needed
voltage_v, current_a Supply's measured output in volts and amperes
output_on Queried output-enable state, 0 or 1; not proof of delivered power
commanded_cathode_mv Software's requested DAC command, captured before the poll; not measured DAC voltage
plasma_target_a Software's PID target, captured before the poll; zero when off
status ok, unavailable or dummy for a poll; output_on, output_off, output_on_failed or output_off_failed for the row a press leaves
error Failure description for an unavailable poll
identity Supply identity on successful polls

Every row is flushed. Existing telemetry files are never overwritten. A file write failure stops this recorder with an explicit application-log message; it does not stop the ADC or alter outputs. Voltage, current and output state are queried sequentially, not simultaneously; the time column marks receipt, not an instrument-provided measurement timestamp. Align with the ADC by time, converting the offset-aware timestamp to the rig's local time as needed. The DAC can clamp its command; the requested value is not proof of the applied voltage. There is one row per attempted poll, not one row per nominal sampling slot: use timestamps and status transitions to measure outage duration, rather than counting unavailable rows as 0.5-second intervals.

On LAN failure: the entire poll is unavailable, its measurement/output fields are empty, and the application Log says Kikusui telemetry LOST once per outage. The background recorder retries and logs BACK on recovery. There is no reuse of the last good reading as a new measurement, no zero substitution, and no change to manual drive. Reconnection verifies model and serial identity before anything is sent, writes included. Stop interrupts a connected read and all retry/sampling waits; hardware shutdown takes precedence over waiting for this recorder. A future PID that depends on telemetry will need its own explicit loss-of-feedback policy.

The existing ADC Ci/Cv columns and the Cathode Measured cell are still the old analog placeholders: they are not these LAN measurements. The separate file preserves the original ADC schema and independent timing. From 4.16.0, the Qt Cathode panel shows Kikusui V/I and recording/output state beneath the manual controls. Hover for the file path and sample receipt time. The display updates every 500 ms from a locked copy of a flushed CSV row. Failed reads clear the numbers on the next display update; readings older than the configured interval plus timeout plus 0.5 s are withheld as stale. Stopping clears the readout. Simulated data is labelled SIMULATED. From 4.17.0 the WebUI has full filament voltage/current cards under Kikusui in Operate, Observe and Monitor. The existing small/big switch sizes these readings too, and folding keeps both values on the summary. The manual control no longer displays the unwired analog Cv placeholder as a measured supply voltage. The ADC columns retain their original meaning.

/api/state carries a separate kikusui object: status, measured voltage_v, current_a, output_on, receipt date, age_s, stale_after_s, query timing, identity and the telemetry filename when available. The main thread publishes the recorder snapshot; the web thread reads plain locked values. Age continues increasing if the main thread stalls, independently of ADC freshness. Non-fresh states omit V/I/output fields, and browser-side expiry also clears the cards if requests hang. These are instantaneous receipt-time measurements, not averaged ADC samples. Filament resistance V/I and power VI can be derived during review; avoid dividing near zero current and compare resistance under similar thermal and operating conditions before interpreting it as thinning.

The two writes

Owner decision 2026-09-15, live, in front of the standing rule of 2026-09-14 that this link was read-only: "Kikusui has LAN now, I want the one-off signal button in our control. Then we turn that one off no matter the dac voltage" and, the same day, "Off and on. Why not? Then we have full cathode control when powered". The link may now send exactly two commands, OUTP 0 and OUTP 1, and nothing else ever: no voltage or current setpoint, no *RST, no status clear. The allowlist in ReadOnlyClient is the whole of it — six words, four queries and these two — and a test holds it to those six and proves that OUTP ON, VOLT, CURR and *RST raise before a byte is sent.

The output is not the drive. The cathode DAC feeds the supply's external current-control input; the output is the supply's own switch between it and the filament. Dropping the drive to zero is not the same as opening the output, which is the press this exists for.

Off is always allowed; on is gated. Off asks nothing of anybody — no Remote switch, no name, no lab's word, no acquisition — exactly as Stop all outputs asks nothing, and Stop all outputs sends it too. On is a setter: the Remote switch on the rig's screen, the lab's word where the machine holds one, and the operator lock, all as any setpoint. On is refused outright while the telemetry is not fresh (the supply is not answering; turn it on at the supply), so nothing is closed onto a filament that nobody can see.

Who sends it. A press never runs on the Qt main thread and never waits on the LAN there. While a run is recording, the press is handed to the recorder's own thread, which carries it between polls on the socket it already owns — one thread, one socket, and a press that arrives mid-poll waits for the poll. With no recorder running, a short daemon thread opens its own connection, verifies the supply's identity, writes, reads back and closes. Either way the OUTP? readback is queried on the same connection and the Log says one of three things:

Kikusui output OFF (confirmed)                   the supply answered 0
Kikusui output OFF sent, readback 1              the write went out; the supply disagrees
Kikusui output OFF FAILED: <reason>; manual drive unchanged

with ON in place of OFF for the other direction, and a fourth line — Kikusui output OFF SIMULATED: no supply is connected — on dummy hardware, where the press moves a simulated flag and nothing else.

In the sidecar. A press writes one row of its own, with the readback in output_on, no voltage or current (the supply was switched, not measured), and status one of output_on, output_off, output_on_failed, output_off_failed. No column is added and the schema line stays controlunit-kikusui/v1; a reader that selects status == "ok" rows is unaffected, and one that counts states should know these four. The recorder publishes that row as its snapshot, so the Live page's Cathode card reads output off sent for the one poll between the write and the next measurement rather than a number nobody measured.

On the web. POST /api/cathode-output takes {"on": false} or {"on": true}. On the Live page the Cathode card's heading line carries a lamp: green while the telemetry is fresh and the output is closed, grey while it is fresh and open, and an uncoloured ring while there is no fresh answer. The word for that state — output on, output off, or the reason there is none — is on the lamp's tooltip and, in full, on the two cathode readout cards; the heading row has no width for it beside the PID/Manual switch, and saying it twice on one screen is the defect fleet's Teaching law names. Pressing the lamp sends the opposite of what it shows; from the dim state it sends only off, which is the press no gate stands in front of.

Transport

Socket to port 5025 (SCPI-RAW), one line per command terminated with LF. The recorder's complete wire command set is:

*IDN?       -> identity (on connection/reconnection)
MEAS:VOLT?  -> output voltage in volts
MEAS:CURR?  -> output current in amperes
OUTP?       -> output-enable state, 0 or 1
OUTP 0      -> open the output   (a press, never the recorder's own doing)
OUTP 1      -> close the output  (a press, gated; see The two writes)

The recorder maintains one socket, sends one query at a time and reads a complete newline-terminated reply before sending another. A single recv is not guaranteed to contain even one complete reply. One timeout bounds the whole poll, including fragmented responses; invalid numbers, output states or identity are failures rather than plausible-looking measurements.

It does not send SYST:COMM:RLST REM, any voltage or current setpoint, reset or clear-status command. Reading must not take over the front panel or clear an instrument alarm. The existing J1 current control remains the actuator: the two writes switch the supply's output, they never set what it delivers.

On 2026-09-14 the connected PWR401L (firmware VER01.24 BLD0056) answered these queries with output off. Ten voltage/current pairs from the office PC took 4.58 ms median, 19.53 ms maximum. This proves the transport, not its reliability during plasma; discharges and bakes are the next evidence.

Six further read-only polls from the Pi itself on the same day succeeded with output off, taking 2.772–3.266 ms per voltage/current/output poll.

First run after deployment

With the machine-local configuration above present, start acquisition from the rig GUI. The Cathode panel should show Recording, V/I and output state; the Log names kikusui_<run timestamp>.csv alongside the ADC file. During normal manual operation, compare those readings with the supply's display. Stop acquisition normally and inspect the paired files. Capture discharges and bakes before setting filament-condition thresholds or changing PID behaviour. The logger sets nothing; the only thing it can switch is the output, and only when somebody presses for it.

Sources