The canonical source is controlunit/settings.yml — the developer's deliberate in-code data dictionary, written as a memory aid after long gaps in work. Channel names, ADC addresses, gains, descriptions, and conversion function identifiers all live there.

"I did that to not remember the modules. So I DID DOCUMENT THERE INTENTIONALLY." — Arseniy

Hardware documentation (board layouts, signal conditioning, pin numberings, datasheets) lives in the external hub: aklab-howto. Do not expect hardware build notes here.

ADC signal channels

Name Signal Sensor / notes
Ip Plasma current Hall-effect sensor on channel 0, on the preanode supply's wire (owner statement 2026-10-07: "plasma current IS preanode current presently"): 25 * (Ip / Vhall - 0.4961) A; supply-corrected since 4.22.0, with the existing provisional 5 A/V sensitivity at 5 V. Against that supply's panel it reads about 0.70 (0.60 A for 0.9 A on 2026-10-02, 1.54 for 2.2 A on 2026-10-05, a flat 3.53 A where the supply limited at 5 A on 2026-10-06), so the true sensitivity is nearer 7 A/V until the known-current calibration says the number. The resting ratio was the ideal 0.5 until 4.22.2 (files up to 2026-10-02 read 0.098 A low); each file's header carries the formula it was written with.
Vhall Hall supply reference Channel 1, read immediately after Ip, 10 V range (the 5 V range clips at 5.0176 V). Raw volts retained for diagnosis; no additional display or curve.
Pu Upstream pressure Pfeiffer single gauge PKR251 (currently operating in Pirani mode — cold-cathode discharge not igniting)
Pd Downstream pressure Ionization gauge, with Place: "downstream" in the settings for the Control dock and the web Gauges card to show beside its short name. Its controller's mode (0 Torr linear, 1 Pa log, or 2 off — the gauge switched off at its controller and declared so on the Control dock or the web Ion gauges card, when the raw volts are kept and Pd_c is NaN) and exponent are recorded per row as IGmode and IGscale, the columns the file has always had.
Pu2 Upstream pressure Second ionization gauge, wired to channel 16 on 2026-09-15, with its own selector on the Control dock and the web Control tab, and Place: "upstream" in the settings for both to show beside its short name. Its mode (the same three codes as Pd's) and exponent are recorded per row as IGmode_Pu2 and IGscale_Pu2, appended after the original columns so nothing an old reader counts on moves.
Bu Upstream Baratron MKS 627, FS = 1 Torr
Bd Downstream Baratron MKS 628B, FS = 0.1 Torr
MFC1 H₂ flow 20 SCCM range
MFC2 O₂ flow 10 SCCM range
Ci Cathode current Channel prepared in ADC map and on PCB; never actually measured. Conversion function exists for consistency.
Cv Cathode voltage Same as Ci — prepared but not deployed
T Membrane temperature Now read off-Pi (NI on Windows). Channel definition remains.
QMS_signal Sync trigger state Boolean — logs whether GPIO sync is active during each row

Hall supply correction (4.22.0)

The reference must measure the supply feeding the Hall sensor, against the same ADC ground. Queezz connected the PSU voltage on 2026-09-30. The prior board inspection identifies the neighbouring terminal as channel 1 beside Ip on channel 0; verify those voltages on the rig before deployment.

Ip names Supply Channel, Nominal Supply, Zero Ratio, and Amperes Per Volt in settings. The values are 5 V, 0.4961 and 5 A/V; the ratio is where the sensor was measured to rest with no current on 2026-09-30 and 2026-10-02, the same within 4 mA across both days. This cancels proportional supply movement in both the zero and the sensitivity; it does not replace a known-current calibration or correct ground offsets. The recorded Ip_c, existing current readouts, and PID use the same correction. Raw Ip and Vhall remain volts, and the CSV header records the formula. Readers must use named CSV columns, as the new signal changes column positions. Slow sampling converts the paired means stored in the row.

A missing, non-finite or out-of-range supply (configured limits 4–6 V) makes Ip_c unavailable (NaN); the event is logged once and recovery is logged too. No stale reference or fixed-voltage fallback is substituted. The PID holds its existing command during that interval. Settings without a supply channel retain 5 * (Ip - 2.52), with optional Zero Voltage and Amperes Per Volt overrides. Settings version 1.5 activates the new canonical map; older local settings need their customizations merged before reuse. The rig had no local settings file on 2026-09-30.

The noise investigation explains why the supply was added. Filters and longer baseline windows are separate work; reassess noise with the newly wired reference first.

DAC outputs

Device Signal Notes
DAC8532 ch1 MFC1 (H₂) setpoint voltage
DAC8532 ch2 MFC2 (O₂) setpoint voltage
MCP4725 Plasma current (Ip) DAC Behind galvanic I²C isolator since Apr 2026

Conversion functions

Conversion polynomials are defined in controlunit/devices/conversions.py with LaTeX in docstrings — physics inline with the code that uses it.

The polynomials come from external datasheets. The PKR251 datasheet, for example, lives in aklab-howto.

Dispatch is settings-driven: each channel in settings.yml carries a Conversion Function key that maps to a function in conversions.py via AdcChannelProps.set_conversion_function(). Adding a new instance of an existing sensor type means editing YAML, not code.

Settings versioning

The repo carries controlunit/settings.yml as canonical schema. The live rig runs ~/.controlunit/settings.yml with its own channel map. A Settings Version key prevents silent drift after a schema bump:

# controlunit/readsettings.py
if local_config["Settings Version"] == config["Settings Version"]:
    config = local_config